Turbidity Sensor OLTU610

Featuring a built-in OLED screen and 3 operation buttons, this compact instrument uses a 660nm laser scattering principle to perform ultra-low turbidity detection with high-resolution down to a 0.001-100 NTU span.

  • Measuring Range: Turbidity: 0.001~100NTU TEMP: 0~45℃
  • Delivery time: 2-5 weeks
  • MOQ: 1
  • Price: Contact Us
Option
  • Anti-twist cap
  • PG13.5
Get A Quote
  • Fast And Safe Logistics
  • Reliable Quality, Guaranteed Performance.

Standard

Model OLTU610
Principle 660nm laser scattering principle
Measuring Range Turbidity: 0.001~100NTU   TEMP: 0~45℃
Resolution Turbidity: 0.001NTU   TEMP: 0.1℃
Accuracy

0.001~40NTU: ±2% of reading or ±0.015NTU, whichever is greater;

40~100NTU: ±5% of reading
Communication RS485 MODBUS-RTU
Display OLED screen
Operation Panel 3 operation buttons
Related Compensation Temperature measurement
Response Time Step response, initial response time < 60 seconds
Calibration Method Multi-point calibration of standard solutions; offset correction, factor correction
Sample Pressure Ensure that the sample pressure before the injection port is 1~2Bar
Sample Flow Rate

100 to 700 mL/min; Ideal flow rate: 200 to 300 mL/min

Working Temp 0~45°C (no freezing)
Storage Temp -10~60℃
Main Material POM, nylon, ABS, stainless steel
Dimension 140mm x 140mm x 330mm
Mounting Main body wall-mounted
Lead Interface M12 connection
Protection Grade IP65
Power Supply 9~18V DC
Power Consumption 2.5W
Cable Standard 2 meters (@can be customized)